Now, using a strontium source and putting it above of the silicon detector with a metal shape to hold it, I'll take new measures.
I turn on the Alibava, the Agilent (photomultiplier) and the Keithley (silicon detector).
Settings:
Agilent - +6V ---> 0.850V Keithley - Cmpl ---> Fixed on 10μA Alibava - 100000 Thrs for each voltage, with Thr in -30.
+25v --> 12V Vrsc ----> I made 8 measures with different values: I saved the files on /tmp/ and I wrote the name like this:
20V, 40V, 60V, 80V, 100V, 120V, 140V and 160V Ex: /tmp/V80-μA0.55-LV0.850-HV12-Thr-30
Where: V= Keithley's voltage
μA= Relative amp to the Keithley's voltage
LV= Agilent's Low Voltage (+6V)
HV= Agilent's High Voltage (+25V)
Thr= Alibava's Thrigger
On V160, I'd problems. I tried 3 times, and Alibava stop working at 40000/50000 Thrs. The same occured on V20.
I put 30000 Thrs, instead of 100000, and it works.
Hans asked me to put insulating tape on some cabes of the Keithley and Agilent, and I did it.
|