Last week, I made measures with the silicon detector and the photomultiplier, but I saved the data on tmp folder...
I lost all the datas of last week.. Now, I'm doing the measures again. The information and details are the same, then I'll just copy here.
The only change will be the folder, changing tmp for Data_analyses.
Now, using a strontium source and putting it above of the silicon detector with a metal shape to hold it, I'll take new measures.
I turn on the Alibava, the Agilent (photomultiplier) and the Keithley (silicon detector).
Settings:
Agilent - +6V ---> 0.850V Keithley - Cmpl ---> Fixed on 10μA Alibava - 100000 Thrs for each voltage, with Thr in -30.
+25v --> 12V Vrsc ----> I made 8 measures with different values: I saved the files on /tmp/ and I wrote the name like this:
20V, 40V, 60V, 80V, 100V, 120V, 140V and 160V Ex: /Data_analyses/11-04-1994/V20-uA0.39-LV0.850-HV12-Thr-30
Where: V= Keithley's voltage
μA= Relative amp to the Keithley's voltage at the end of the measure
LV= Agilent's Low Voltage (+6V)
HV= Agilent's High Voltage (+25V)
Thr= Alibava's Thrigger
|